Model Number 209999 |
Device Problem
Material Frayed (1262)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Event Description
|
Boot has chip in it which is sharp.Case type: tka.
|
|
Manufacturer Narrative
|
Additional information: update to catalog number and gtin.Boot has chip in it which is sharp.Case type: tka.Product evaluation and results: the boot assembly showed a broken edge with carbon fiber splinters.Product history review: review of the device history records indicate (b)(4) were manufactured and accepted into final stock on 06-14-2017 with no reported discrepancies.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 210080, lot 201743060801 shows 3 additional complaints related to the failure in this investigation.(b)(4).Conclusions: per d03391, preventive maintenance is where an action occurs that identifies device deterioration which may compromise function.Under pm conditions no patient was involved and no actual or potential patient harm existed for the alleged event.The failure was confirmed via visual inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
|
Event Description
|
Boot has chip in it which is sharp.Case type: tka.
|
|