Qn#(b)(4).The sample was returned for evaluation.Upon receipt the unit was visually inspected for outward, visible signs of misuse/abuse/neglect.Nothing of any significance was noted in terms of physical damage.The unit was inspected for cracks, wire frays, plastic case cracks/fractures, burn areas/wires, damaged power cord strain reliefs, and nothing was noted.No significant amount of lint built up on fan or cooling ports.The unit was gently rotated and then lightly shaken to determine if any loose components might be moving around inside the plastic case.Nothing loose was noted.The neptune was then connected to 110 vac.The unit passed the initial power connect test.The unit also navigated through the power-on self-test (p.O.S.T.) with no issues.The next test was the temperature display accuracy test.The neptune displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for the functional bench test where water, an adult breathing circuit (880-36kit), 10 lpm of air pressure, and a 382-10 concha smart water column was connected to the unit for a real time operational scenario.Temperature was set at 37 c, mode was invasive, full rainout on the moisture scale.The concha smart column was placed into the neptune before the bench test was started.The neptune was turned on, values set.The column was empty.The neptune low water indicator lamp came on.Water was added to the column.The unit successfully negotiated all pre-operational self-tests again and was functioning real time.The unit functioned without any interruption or functional anomalies for ~2.5 hours.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
|