• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problem Overfill (2404)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/10/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint mr290v humidification chamber was returned to fisher & paykel healthcare (f&p) in (b)(4) for investigation.Visual inspection and water level testing of the returned mr290 chamber was performed.Results: visual inspection revealed no damage to the chamber.The floats were found to be moving freely.During the functional test, the water level of both primary and secondary floats was checked.Both floats operated correctly, and the water level attained was within specification.Conclusion: we were unable to determine the cause of the reported issue, as the complaint mr290 chamber was found to be operating correctly.Overfilling of water is usually caused by both the primary and secondary float mechanisms in the mr290 chamber being disabled.In circumstances where the primary float is disabled, the water may overfill above the black water level line of the chamber but the secondary float will serve as a backup to prevent the water from overflowing outside the chamber port and entering into the breathing circuit.All mr290 chambers have float function and valve testing performed twice during production.A failure of any test would result in rejection of the chamber before distribution.The user instructions that accompany the mr290 humidification chamber specify in the warning section "do not use the chamber if the water rises above the maximum water level line" along with a diagram directing the user to check the water level on the chamber.A written description of the meaning of the symbols used is also included in the user instructions.It also states the following: "set appropriate ventilator alarm." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.".
 
Event Description
A distributor in (b)(6) reported on behalf of a healthcare facility to a fisher & paykel healthcare (f&p) field representative that an mr290v vented autofeed humidification chamber was overfilled immediately after starting to use.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology dr.
suite 100
irvine, CA 92618
9494534000
MDR Report Key9528689
MDR Text Key181959172
Report Number9611451-2019-00972
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number190307
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2019
Initial Date FDA Received12/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-