Device Problem Microbial Contamination of Device (2303)
Patient Problem Unspecified Infection (1930)
Event Date 11/29/2019
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chimaera.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(4).Livanova initiated an investigation a review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with environmental mycobacterium.The customer required deep disinfection procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key9530254
MDR Text Key177944848
Report Number9611109-2019-01014
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received12/30/2019
Type of Device Usage N
Removal/Correction NumberZ-2076/2081-2015