ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2019
Event Type  malfunction  
Manufacturer Narrative
The reporter's test strips were provided for investigation where they were tested using a retention meter and controls.Testing results (qc range = 2.5 - 3.7 inr): qc 1: 3.0 inr, qc 2: 3.0 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.Customer stated they were taking antibiotics.Per product labeling, when other medications, like antibiotics, are taken simultaneously it can lead to either an increase or a decrease in prothrombin time (inr).Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." relevant retention test strips (lot 397398) were tested in comparison with the master lot coaguchek xs pt.For this purpose two human blood samples from marcumar donors and two internal reference meters were used.Retention samples were acceptable.No error messages occurred.
 
Event Description
The initial reporter received a questionable result from coaguchek vantus serial number (b)(4).The result from the coaguchek vantus meter was 3.0 inr.The result from a coaguchek pro meter was 3.1 inr.The result from a laboratory using an unknown reagent was 2.2 inr.Refer to the medwatch with patient identifier (b)(6) for the coaguchek pro meter.The therapeutic range was 2.5-3.5 inr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key9530900
MDR Text Key217948302
Report Number1823260-2019-04581
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2020
Device Catalogue Number04625374160
Device Lot Number39739821
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2019
Initial Date FDA Received12/30/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age42 YR
Patient Weight81
-
-