Additional information was added.The device was received for evaluation.A visual inspection was performed, and it was noted that there was a connection issue due to the collar of the male luer being glued with solvent.The reported condition was verified.The cause of the reported condition was excess of solvent on the male luer ¿ tubing joint due to a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|