AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC Back to Search Results
Model Number 8360-10
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2019
Event Type  malfunction  
Manufacturer Narrative
Manufacturing evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report, if applicable.
 
Event Description
It was reported that there was an issue with the prestige grasper.In a robotic uterine myomectomy procedure, the tip broke during positioning.It occurred after introduction of the instrument into the trocar.The jaw broke off; the pieces were retrieved and did not fall into the patient's abdomen.The surgery was completed successfully with another device.There was no patient harm or intervention required.
 
Manufacturer Narrative
D section: batch # updated; supplier confirmed as (b)(4).Manufacturing evaluation: additional information / investigation results will be provided in a supplemental report, if applicable.
 
Manufacturer Narrative
Additional information - d4 (udi), h7 corrected information - d4 (lot #) manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.An investigation of the device manufacturing records was conducted by the manufacturer for the lot # of the device in question.No non-conformances were reported.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Additionally, historical scrap rates were reviewed with no increase observed in scrap related to the complaint issue.Previous investigations performed against devices returned with distal braze failures identified improvement opportunities following a review of the tube sub assembly test procedure work instruction (wi), the brazing procedure wi, and the brazed joint buffing wi.While the tube subassembly joint is 100% percent tested with a torsional force, there was no requirement for applying a bending force to the joint.Therefore, a manual bend test was added to the wi.Additionally, a review of the torque test fixture and accompanying wi, noted the potential for the tube to slip inside the collet during inspections allowing for a defective part to potentially pass this test.The tube sub assembly test procedure was further updated to note this potential failure mode and to define the process for cleaning the parts and fixture/collet with alcohol prior to use.A review of the brazing procedure wi revealed that the glass tube was too short to effectively seal the brazing area off from the surrounding environment.Without a proper seal the brazing area could have insufficient argon present to facilitate effective brazing.The brazing procedure wi was updated to include a check for this condition prior to brazing.Additionally, the supplier updated the wi to optimize the order of operations of when flux is applied, the soldering ring is assembled, and the tube is loaded.This change ensured that flux would be present throughout the entire joint space and allow for proper solder travel.Furthermore, a functional review and visual examination of the nest, which the tube sub assembly sits into, was performed.This review revealed that the two argon access holes were clogged.Therefore, the associated preventative maintenance activities were updated to monitor the access holes and prevent a recurrence of buildup.Finally, the supplier updated the brazed joint buffing wi to note the potential failure mode of excessive buffing, which could remove too much material and weaken the joint.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.However, based on prior evaluations of complaint devices reported with a failure mode of distal braze break/separation, this event likely occurred due to inadequacies in the defined production process which limited the device performance.Aesculap inc.Opened a corrective action/preventive action (capa) for further evaluation of the design transfer of this device.
 
Event Description
No update required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE ATRA GRASPER DBL-ACT 5MM
Type of Device
LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
lindsay chromiak
3773 corporate parkway
center valley, PA 18034
6109849072
MDR Report Key9532043
MDR Text Key173287655
Report Number2916714-2019-00143
Device Sequence Number1
Product Code NWV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8360-10
Device Catalogue Number8360-10
Device Lot NumberL50510290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received01/08/2020
11/02/2021
Supplement Dates FDA Received03/30/2020
11/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-