• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 5000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 5000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938743
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 5000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was leaking from the spike port.The leak was discovered after filling the bag with fluid using the needle port, prior to patient use.The reporter stated that the spike port had not been accessed.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured from august 04, 2018 - august 05, 2018.The device was received for evaluation.Visual inspection was performed which did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak at the spike port cap from the spike port.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9532173
MDR Text Key173096537
Report Number1416980-2019-07246
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477275
UDI-Public(01)00085412477275
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2021
Device Catalogue NumberH938743
Device Lot Number60142546
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received01/20/2020
Supplement Dates FDA Received01/30/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-