Catalog Number H938738
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 500ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was leaking from the middle spike port.The leak was discovered prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was manufactured from april 30, 2019 - may 02, 2019.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to (update to evaluation method code and evaluation result code).The actual device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed which revealed spike port cap leakage at the base of the spike port tubing.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9532208
MDR Text Key173096145
Report Number1416980-2019-07247
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477220
UDI-Public(01)00085412477220
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2022
Device Catalogue NumberH938738
Device Lot Number60182409
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2020
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received01/29/2020
02/11/2020
Supplement Dates FDA Received01/31/2020
02/17/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-