Model Number 10014914 |
Device Problem
Failure to Prime (1492)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
It was reported that the tubing failed to prime.It was not reported that patient care was delayed and it did not contribute to, or result in serious adverse impact to patient.
|
|
Manufacturer Narrative
|
Correction; e.1, e.2, and g.3.The customer¿s report that the tubing failed to prime was confirmed based on functional testing.Functional testing observed that fluid was not able to flow due to excess solvent at the engagement between the tubing and male luer components.The set sample was visually inspected for kinks, holes/tears in the tubing or damages to the components.The set was attempted to be re-primed using a blue dyed fluid filled lab syringe.It was observed there was resistance and the fluid would not flow past the engagement of the micro-bore tubing and male luer components.The tubing area directly above the male luer was manually cut.The fluid was now able to exit the cut tubing end.Visual inspection under magnification through the tubing to the male luer observed an occlusion identified as excess solvent.The root cause was identified as a manufacturing issue of excess solvent at the tubing engagement.
|
|
Event Description
|
It was reported that the tubing failed to prime.It was not reported that patient care was delayed and it did not contribute to, or result in serious adverse impact to patient.
|
|
Search Alerts/Recalls
|