The lot was manufactured from october 12, 2018 - october 16, 2018.H10: the device was received for evaluation.Visual inspection revealed leak/backflow at the fill port after the fill port cap was removed from the port.Additional examination on the device revealed the cause of leak/backflow at the fill port was due to a glass fragment lodged under the device checkband.The reported condition was verified.The cause of the glass fragment could not be determined, however the most probable cause is due to user filling techniques related to glass ampule usage.No manufacturing process step would allow glass fragments to be introduced near or adjacent to the fill port.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|