H4: the device was manufactured from july 10, 2019 - july 12, 2019.H10: the actual device was received for evaluation.A visual inspection was performed using the naked eye which revealed traces of silicone oil located on the interior wall of the housing.Silicone oil is used during the manufacturing process and therefore, is not considered a particulate or foreign matter.Occasionally, silicone oil can drip on the interior wall of the housing; this condition is cosmetic and has no impact on the functionality nor safety of the product.Therefore, the reported condition of particulate matter was not verified.A functional testing was not performed for this complaint.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|