The explanted devices were returned for product analysis.Product analysis was performed on the explanted lead and generator.Visual analysis confirmed quadfilar coil appeared to be broken in two locations not related to explant of the device.Pitting was observed preventing the identification of the coil fracture type in the first location.In the second location, the coil was worn to the point of fracture with flat spots on the coil.Lead insulation was observed to be abraded due to wear.Based on the findings in the product analysis, there is evidence to suggest a discontinuity in the returned portions of the device.The electrode array section was not returned for analysis, and therefore an evaluation and resulting commentary cannot be made on that portion of the lead.No additional relevant information has been received to date.
|