• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Date 10/29/2019
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2 implantable collamer lens, -13.00/3.50/92 (sphere/cylinder/axis) into the patient's right eye (od) on (b)(6) 2019.On (b)(6) 2019 the lens was exchanged for a shorter length lens due to excessive vault.This exchange resolved the problem.
 
Manufacturer Narrative
Additional information: h3- device evaluation: lens returned in a microcentrifuge vial with moisture on lens and clear surgical residue on product.Visual inspection found no visible damage to lens and residue on lens.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key9533416
MDR Text Key173252039
Report Number2023826-2019-02484
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received01/08/2020
Supplement Dates FDA Received01/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age20 YR
-
-