BAXTER HEALTHCARE CORPORATION 500 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938738
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the ¿heat seal¿of a 500 ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag ¿was not done properly creating holes outside the bag¿.This was identified prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was manufactured from april 10, 2019 - april 11, 2019.The device was received for evaluation.Unaided visual inspection was performed which observed a detached top port tubing from the inside of the bag and protruding outside the back of the bag.Functional testing was performed which revealed the top of the fill port leaking outside the back of the bag when attempting to fill the bag.The reported condition was verified.The cause of the condition was due to a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9533473
MDR Text Key173201807
Report Number1416980-2019-07279
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477220
UDI-Public(01)00085412477220
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Catalogue NumberH938738
Device Lot Number60179094
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received01/09/2020
Supplement Dates FDA Received01/31/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-