ENCORE MEDICAL L.P. RSP; RSP BASEPLATE, 30MM, W/P2 COATING Back to Search Results
Model Number 506-03-138
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 12/04/2019
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Second revision surgery - center screw on baseplate and one peripheral screw broke.
 
Manufacturer Narrative
The reason for this revision surgery was reported as broken screw.The previous surgery and the surgery detailed in this event occurred 1 year and 11 months apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports (ncmr) associated with the product that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to broken screw.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical are soft tissue impingement, excessive loads or torques, poor bone quality, patient activities and trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
RSP BASEPLATE, 30MM, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key9535312
MDR Text Key173296408
Report Number1644408-2019-01273
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024594
UDI-Public(01)00888912024594
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/08/2023
Device Model Number506-03-138
Device Catalogue Number508-32-204
Device Lot Number769P1343
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/31/2019
Supplement Dates Manufacturer Received01/08/2020
Supplement Dates FDA Received02/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-