ZOLL CIRCULATION AUTOPULSE® PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2019
Event Type  malfunction  
Manufacturer Narrative
The report of the autopulse platform (sn (b)(4)) displaying a user advisory (ua) 45 (driveshaft not at "home" position after power-on/restart) error message was confirmed during functional testing and confirmed during archive data review.The investigation findings revealed that the root cause was due to the driveshaft not to be at home position in which likely attributed to user error.Visual inspection of the returned platform was performed and found no physical damage.The archive data was reviewed and contained multiple user advisory (ua) 45 (driveshaft not at "home" position after power-on/restart) error message occurred on the reported event date.Evaluation of the platform during initial power up, revealed a ua 45 error message on the user control panel.It was indicated that the driveshaft was not in home position.The driveshaft was readjusted back to home position.The autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test battery until discharged without any fault or error.The platform passed all functional tests and is ready for clinical use.Note ua 45 alerts the operator that the autopulse driveshaft is not at its home position when the platform was powered on.This user advisory will persist until the driveshaft is returned to its home position.Per the autopulse user guide instruct, to clear ua 45, the operator needs to pull up the lifeband until the chest bands are full extended.This action will move the driveshaft to its home position.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
During patient use, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 45 (driveshaft not at "home" position after power-on/restart) error message on the user control panel.The crew was unable to clear ua 45 error message.The crew immediately performed manual cpr.No known impact or patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key9537002
MDR Text Key179429685
Report Number3010617000-2019-01202
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/10/2019
Initial Date FDA Received12/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-