The customer reported that "on (b)(6) 2019, the nursing staff found a cyanotic patient who had an asystole and was connected to the monitor.Neither an acoustic nor a visual alarm was triggered on the monitor.Resuscitation measures began immediately, the patient died.
|
H3 and h6: a philips field service engineer (fse) went onsite to collect and review the audit log for the time of the reported event.During the onsite visit, a nurse reported that the ecg/arrhythmia alarms had been turned off by a user, though it was unknown why that was done.This was confirmed during the analysis of the audit log.The ecg/arrhythmia alarms were deactivated for bed mits3 on (b)(6) 2019 at 17:22:41 and were not switched back on until after the reported death, on (b)(6) 2019 at 05:48:00, which resulted in the fact that no ecg/arrhythmia alarms were announced during this time.This delayed the response to the patient.The customer was advised via email and letter of the investigation outcome.Additionally, onsite training for the clinical staff was requested by the customer with the regard to alarm settings/behavior.The monitor worked as intended and there was no malfunction of the device.This issue was caused by the fact that the ecg/arrhythmia alarms were switched off by the user at the reported time.The available information from this report does not support that this failure represents a systemic, design, or labeling problem.No further investigation or action is warranted.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|