(b)(4).Corrected data: section d.1.Brand name corrected to lma proseal reu introducer (3,4,5).Section d.4.Catalog# corrected to 151030.The sample was returned for evaluation.The manufacturer reports a visual exam was performed and it was observed that the blue silicone material was damaged and was detached from the metal handle.Upon closer observation it was noticed there were scratches across the stainless steel made introducer body.It was also found that the bond of the insert-molded silicone strip was damaged.Based on the investigation performed the reported complaint was confirmed.The manufacturer states that scratches/damage on the bond of the insert-molded silicone strip will adversely impact the bonding of the silicone strip on the metal made proseal introducer and with add on pushing force at the silicone tip, the properties of the bond will be further damaged causing failure of the product.The device is reusable and will be worn out depending on usage volume and handling methods.
|