U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL SILICONE TRICOT 25BX; STATLOCK FOLEY SWIVEL SILICONE TRICOT 5ST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL SILICONE TRICOT 25BX; STATLOCK FOLEY SWIVEL SILICONE TRICOT 5ST Back to Search Results
Catalog Number FOL0102
Device Problems Disconnection (1171); Defective Component (2292)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the clip on the foley statlock was defective.It was later reported from the ibc via email 20dec2019 that the clip part of the stat lock became disconnected from the adhesive part of the device 3 weeks of use.
 
Event Description
It was reported that the clip on the foley statlock was defective.It was later reported from the ibc via email 20dec2019 that the clip part of the stat lock became disconnected from the adhesive part of the device 3 weeks into use.
 
Manufacturer Narrative
The reported event was confirmed.Visual inspection noted one foley statlock received with the original packaging.Visual evaluation also noted that the clamp was disconnected from the swivel base on return.This failed to meet specifications per the standard stating, the device will be rejected due to damaged component or assembly.Although the reported event was confirmed, a root cause could not be determined.A potential root cause for this failure could be incorrect dimensions between mating parts.The product was influenced by the reported failure.The product was used for treatment purposes.The product failed to meet specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿prep.1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 2.5 cm of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel.7.Align the statlock® stabilization device over securement site leaving 2.5 cm of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique disengage.1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away from skin." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL SILICONE TRICOT 25BX
Type of Device
STATLOCK FOLEY SWIVEL SILICONE TRICOT 5ST
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9545178
MDR Text Key178645738
Report Number1018233-2020-00080
Device Sequence Number1
Product Code EYJ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 01/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2021
Device Catalogue NumberFOL0102
Device Lot NumberJUCYF144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2020
Initial Date Manufacturer Received 12/13/2019
Initial Date FDA Received01/03/2020
Supplement Dates Manufacturer Received01/28/2020
Supplement Dates FDA Received01/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-