It was reported that this implantable cardioverter defibrillator (icd) declared code 1004 which is indicative of shorted circuit condition detected during shock delivery.The right ventricular (rv) lead is a non-boston scientific lead.The rv lead exhibited noise and oversensing that resulted in delivery of inappropriate shock therapy.Post inappropriate shock delivery, device therapy was turned off until the generator could surgically be replaced.Noise on the rv lead was able to be replicated through isometric manipulation movements.An invasive procedure was performed.The icd was replaced and the rv lead was surgically abandoned.No additional adverse patient effects were reported.
|