Model Number G310
Device Problems Restricted Flow rate (1248); Overheating of Device (1437); Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g310 scaler, there was intermittent water flow and the handpiece was overheating; no injury resulted.
 
Manufacturer Narrative
Debris buildup in the water supply hose, clogged water filter restricting water flow.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON 300 SERIES ULTRASONIC SCALING SYSTEM UNIT
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key9551102
MDR Text Key186725460
Report Number2424472-2019-00227
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K150535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG310
Device Catalogue NumberG310SCLR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2019
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received01/06/2020
Supplement Dates Manufacturer Received12/05/2019
Supplement Dates FDA Received02/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-