The patient was extremely disappointed with therapy and when interrogating the device, a lead issue came up stating, the "posterior tip to ring circuit open." however, the patient was able to feel a sensation of therapy during a real test set at 7.0m.A.The reshape field specialist adjusted the patient's setting to 4.50m.A.The patient continued to receive therapy following the adjustment.
|
The patient was extremely disappointed with therapy and when interrogating the device, a lead issue came up stating, the "posterior tip to ring circuit open." however, the patient was able to feel a sensation of therapy during a real test set at 7.0m.A.The reshape field specialist adjusted the patient's setting to 4.50m.A.The patient continued to receive therapy following the adjustment.
|