• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2019
Event Type  malfunction  
Manufacturer Narrative
Occupation: ir manager.Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It was reported an ultrathane mac-loc locking loop multipurpose drainage catheter was placed in an unknown patient for drainage.Once the drain was placed, the physician noted "a crack in the hub of the drain." the drain was removed and replaced with a new, similar device.Additional information indicated that the device leaked where the hub meets the catheter.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Concomitant medical product received on: 19dec2020.Investigation-evaluation.It was reported that the hub of the catheter within a ultrathane mac-loc locking loop multipurpose drainage catheter cracked and was leaking.This incident was reported by hospital-univ of pennsylvania, in the united states.The device was removed, and another was placed to successfully complete the procedure.No adverse effects were reported.A review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device, as well as a visual inspection, functional test, and dimensional verification, were conducted during the investigation.The complainant returned one 10.2 fr ult catheter to cook for investigation.Biomatter was present throughout the device.No visible damage to the catheter shaft or the mac-loc hub was noted.The device was leak tested and no leak was detected.Upon further investigation, the mac-loc hub was not found to be cracked.All dimensions deemed relevant to the reported failure were analyzed and confirmed that the device was manufactured within specification.Additionally, a document-based investigation evaluation was performed.The proper procedures are in place to identify and prevent this failure mode prior to device distribution.The risk specifications covering mac-loc drainage catheters include both hub separation and leakage as a potential failure mode.The identified risk controls include manufacturing quality control checks and process validation.The technical files covering mac-loc drainage catheters indicate that the risks associated with these devices are acceptable when weighed against the benefits.The device history record (dhr) was reviewed.The final lot and relevant subassembly lots revealed no related nonconformances.A database search for complaints on the reported lot found no additional complaints reported from the field.At this time, there is no evidence that nonconforming product from this lot exists in house or in the field.The ifu supplied with the mac-loc drainage catheters instruct that the product should be inspected prior to use to ensure no damage has occurred.Based on the information provided, inspection of returned product, and results of the investigation, it was concluded a component failure without a design or manufacturing deficiency contributed to this incident.The appropriate personnel have been notified.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Type of Device
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9566073
MDR Text Key177400212
Report Number1820334-2020-00088
Device Sequence Number1
Product Code GBO
UDI-Device Identifier00827002095021
UDI-Public(01)00827002095021(17)220829(10)9981291
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/29/2022
Device Model NumberN/A
Device Catalogue NumberULT10.2-38-25-P-6S-CLM-RH
Device Lot Number9981291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Initial Date Manufacturer Received 11/21/2019
Initial Date FDA Received01/08/2020
Supplement Dates Manufacturer Received03/04/2020
Supplement Dates FDA Received03/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-