MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D142
Device Problems Device Alarm System (1012); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/29/2019
Event Type  Injury  
Manufacturer Narrative
A thorough evaluation of the device was performed upon receipt at our post market quality assurance laboratory.Review of device memory indicated that a charge timeout alert (> 45 seconds) had been recorded.The device case was opened(, and visual inspection noted that one of the two high voltage (hv) capacitors was slightly swollen).The hv capacitor was analyzed, and it was concluded that the hv capacitor experienced internal arcing caused by a low resistance pathway between anode and cathode plates within the capacitor.This capacitor issue impacted the device's ability to provide shock therapy because the hv capacitors could not be fully charged, but brady pacing, telemetry, and other device functionality remained available.
 
Event Description
It was reported that this device was emitting tones and this patient presented to the emergency room.Interrogation revealed a code 1007 indicating a charge timeout.A boston scientific technical services consultant recommended device replacement.The device was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN EL ICD DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key9570859
MDR Text Key174368199
Report Number2124215-2019-27851
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526534249
UDI-Public00802526534249
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S306
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/26/2020
Device Model NumberD142
Device Catalogue NumberD142
Device Lot Number545845
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2019
Initial Date Manufacturer Received 10/29/2019
Initial Date FDA Received01/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age82 YR
-
-