Model Number 2426-0007 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/17/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.Ethnicity: not hispanic/latino.Race: (b)(6).
|
|
Event Description
|
It was reported that tubing expanded and ruptured while attached to a pressure infuser set at 2.1ml/sec, for a total volume of 78ml of ct scan imaging dye, with a maximum pressure set of 300psi during a ct scan exam.The roller clamp was closed above the distal port of the primary tubing.A 20 gauge iv in the antecubital vein was used for the injection.There was no patient harm.
|
|
Event Description
|
It was reported that tubing expanded and ruptured while attached to a pressure infuser set at 2.1 ml/sec, for a total volume of 78 ml of ct scan imaging dye, max pressure set at 300 psi during a ct scan exam.The roller clamp was closed above the distal port of the primary tubing.A 20 gauge iv in the antecubital vein was used for the injection.It was further confirmed during follow up, that there was no patient harm or impact as a result of this event.
|
|
Manufacturer Narrative
|
The customer¿s report that tubing expanded and ruptured while attached to a pressure infuser set was confirmed.The set was visually inspected for kinks, holes/tears in the tubing or damages to the components.Visual inspection noted a section of the tubing was ruptured and deformed that extended 0.4825inches in length.The affected tubing area was weakened or softened from the damage.No other anomalies were observed.Functional testing could not be performed due to the ruptured tubing.The root cause of the ruptured tubing was due to the excessive pressure of 300 psi that exceeded that maximum pressure that the model 2426-0007¿s tubing can handle.
|
|
Search Alerts/Recalls
|