The patient was undergoing a thrombectomy procedure in the right coronary artery (rca) using an indigo system catrx aspiration catheter (catrx) and a penumbra system aspiration pump max 110 (pump max).During the procedure, the physician was unable to aspirate in the target vessel using the catrx and pump max; therefore, the catrx was removed, and the hospital staff troubleshooted the pump max to determine if there were any air leaks.Upon removal, the catheter was observed to have no noticeable damage, and the pressure gauge on the pump max was reading at the standard -28 inhg.It was reported that the hospital staff did not find any issue with the pump max.The procedure was completed using another catrx, the same pump max, and a stent was placed in the vessel.There was no report of an adverse effect to the patient.
|
Please note that the device associated with this complaint was expected to be returned; however, additional information received from the penumbra sales representative indicated that the device is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|