TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 6382
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2019
Event Type  malfunction  
Event Description
It was reported that during priming of the device for a cardiopulmonary bypass (cpb) procedure, the flow sensor pin came loose, causing the flow sensor to not be able to close.The pin was reinserted, but would fall out again after some time.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
81 evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician observed that the flow sensor did not have a hinge pin, preventing it from working properly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9572362
MDR Text Key193426529
Report Number1828100-2020-00013
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6382
Device Catalogue Number6382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2019
Initial Date FDA Received01/09/2020
Supplement Dates Manufacturer Received01/14/2020
04/23/2020
Supplement Dates FDA Received02/06/2020
05/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-