• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JUGGERKNOT 1.0MM MINI 3-0 NDLS; SCREW, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JUGGERKNOT 1.0MM MINI 3-0 NDLS; SCREW, FIXATION Back to Search Results
Catalog Number 912082
Device Problem Difficult to Advance (2920)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/27/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a finger tendon repair procedure, the juggerknot mini¿s transparent sleeve was stuck during the insertion process.Therefore, the juggerknot suture cannot be deployed in the bone.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, d11, g4, g7, h1, h2, h3, h6, h10 visual evaluation of the returned product identified the clear sleeve was fully deployed upon receipt.The drill bit was not returned.One side of the inserter tip was fractured.The function check was conforming and the device works as intended.Device history record was reviewed and no discrepancies were found.Review of the product determined that the product functions as intended for sleeve jammed issue.However, a definitive root cause for fracture cannot be identified.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERKNOT 1.0MM MINI 3-0 NDLS
Type of Device
SCREW, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9574649
MDR Text Key204241008
Report Number0001825034-2020-00153
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
PMA/PMN Number
K140908
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number912082
Device Lot Number353060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/02/2020
Initial Date FDA Received01/10/2020
Supplement Dates Manufacturer Received06/03/2020
Supplement Dates FDA Received06/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age24 YR
Patient Weight77
-
-