(b)(4).Batch # t93x04.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled, and upon disassembly, the advancer was confirmed to be bent and seven clips were found inside clip track.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.The reported complaint was confirmed.
|