• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET GMBH MAGNUS CARB.-FIBRE TABLE TOP, LONG; TABLE, RADIOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET GMBH MAGNUS CARB.-FIBRE TABLE TOP, LONG; TABLE, RADIOLOGIC Back to Search Results
Model Number 118016A2 - MAGNUS CARB.-FIBRE TABLE TOP, LONG BOLUS
Device Problem Break (1069)
Patient Problem Laceration(s) (1946)
Event Date 01/06/2020
Event Type  Injury  
Manufacturer Narrative
It was reported that the carbon table top was damaged.A getinge ¿ maquet service technician was on site and inspected the product in question.It was detected that edges of the carbon table top were chipped off.This kind of damage is caused due to collision of the table top and obstacles under the table top or when colliding with obstacles against the table top.In the instructions for use (ifu) the user is warned concerning the risks related to collisions and advised as follows: "during the adjustment procedures, always pay attention to the or table and accessories and avoid collisions.Ensure that tubes, cables and drapes are not trapped." besides the user is told in the ifu not to use a defective product.We assume that this issue occurred due to a use error.The product was damaged due to a use error and afterwards the product was used despite the mentioned damages.No serious injury or required medical intervention was reported.Since the extent of the reported cut is unknown, we err on the side of caution and report this as serious injury.Getinge-maquet gmbh provides product failure investigation, analysis and resolution for the device described in this report.The complete event site name is queen elizabeth university hospital glasgow and the postal code is g51 4tf.
 
Event Description
It was reported that the table top was damaged due to a collision.A user has cut his or her hand on the damaged surface during cleaning.Manufacturer reference# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNUS CARB.-FIBRE TABLE TOP, LONG
Type of Device
TABLE, RADIOLOGIC
Manufacturer (Section D)
MAQUET GMBH
kehler strasse 31
rastatt
Manufacturer (Section G)
HOLGER ULLRICH
maquet gmbh
kehler strasse 31,
rastatt 76437
GM   76437
Manufacturer Contact
maquet gmbh
kehler strasse 31,
rastatt 76437
MDR Report Key9574991
MDR Text Key174942767
Report Number8010652-2020-00002
Device Sequence Number1
Product Code KXJ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number118016A2 - MAGNUS CARB.-FIBRE TABLE TOP, LONG BOLUS
Device Catalogue Number118016A2
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/06/2020
Initial Date FDA Received01/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-