• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG 1ST GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG 1ST GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 102954
Device Problems Noise, Audible (3273); Excessive Heating (4030)
Patient Problem Thrombus (2101)
Event Date 08/02/2019
Event Type  Injury  
Manufacturer Narrative
The centrimag motor was captured under mfr# 2916596-2019-03827.The centrimag blood pump was captured under mfr# 2916596-2019-03835.No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
The patient was implanted with an extracorporeal circulatory support pump on (b)(6) 2019.It was reported that the patient was in cardiogenic shock and was initiated on dual centrimag acute support with an oxygenator attached to the right-sided support.Going into the intensive care unit (icu), the left ventricular assist device (lvad) was set at 4000 with 4.7lpm flow and the right vad was set at 2700 with 4lpm flow.Approximately an hour after getting into the icu from the operating room, the cardiologist received an s3 alarm with a loud noise coming from the motor and pump on the lvad and that the motor was hot to the touch.Healthcare professionals immediately switched to the backup motor and there have been no alarms since.The patient was in critical health status due to being in shock and not due to the event.It was also reported that there was additional incident on (b)(6) 2019 in the evening following the switch to full backup system due to rattling sound from pump and s3 alarm.No further alarms have occurred since the switch to full backup system.Photos were provided of the pump secured in the motor on the new system to ensure secure pump placement.On (b)(6) 2019, the patient experienced significant flow fluctuations and apparent suck down at drainage cannula.Volume was given to the patient but flow fluctuations persisted.The patient was taken to the catheterization lab where a "huge left ventricle thrombus" was identified.This was thought to be the cause of the drop in flow and flow fluctuations and the patient was otherwise doing fine.Treatment for the thrombus was to explant the centrimag and implant with a heartmate 3 lvad on (b)(6) 2019.The centrimag console was added as a result of the investigation that concluded other causes may have contributed to the reported events besides thrombus.
 
Manufacturer Narrative
Section d4: the serial number of the centrimag 1st generation primary console was requested but was not provided.Manufacturer's investigation conclusion: the reported event of a s3 alarm was not confirmed.The centrimag 1st generation primary console was not returned for analysis and no additional information regarding the console was provided.The root cause for the reported s3 alarm was unable to be conclusively determined through this analysis.Centrimag motor instructions for use instructs to always have a back-up centrimag motor available.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 1ST GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key9576361
MDR Text Key176011415
Report Number2916596-2020-00153
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K093832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number102954
Device Catalogue Number102954
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2019
Initial Date FDA Received01/10/2020
Supplement Dates Manufacturer Received01/24/2020
Supplement Dates FDA Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age62 YR
Patient Weight111
-
-