• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. INTRALASE; PATIENT INTERFACE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. INTRALASE; PATIENT INTERFACE Back to Search Results
Model Number 590106AN
Device Problem Decrease in Suction (1146)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2019
Event Type  malfunction  
Manufacturer Narrative
Expiration date is unavailable at the time of this report.Manufacturing date is unavailable at the time of this report.The patient interface (pi) suction ring may lose suction during a procedure.Label copy states corneal fixation vacuum loss can occur.There are several factors that may contribute to suction issues such as doctor¿s technique in applying the suction ring to the cornea, doctor¿s technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient¿s cornea and the suction ring.If there is any further relevant information, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
The surgeon reported that he experienced suction loss from the beginning of docking and then the loss of suction moved in during the side cut.The surgeon switched out the patient interface (pi) and re-applied suction and performed a side cut only.The laser application was successful.Both eyes were completed as scheduled.No further medical intervention was necessary.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRALASE
Type of Device
PATIENT INTERFACE
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9576406
MDR Text Key175197870
Report Number3006695864-2020-00046
Device Sequence Number1
Product Code HNO
UDI-Public(01)(10)UNKNOWN
Combination Product (y/n)Y
PMA/PMN Number
K060372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/01/2005,11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number590106AN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 12/19/2019
Initial Date FDA Received01/10/2020
Supplement Dates Manufacturer Received10/25/2020
Supplement Dates FDA Received11/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FEMTOSECOND LASER, SERIAL NUMBER: (B)(6)
Patient Age21 YR
-
-