OLYMPUS MEDICAL SYSTEMS CORP. OES URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P5
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Physical Entrapment (2327); No Information (3190)
Event Date 10/14/2019
Event Type  Injury  
Manufacturer Narrative
The reported scope was not returned to the service enter for evaluation; therefore, the cause of the reported event cannot be determined a this time.As part of our investigation, multiple follow-ups were made to the user facility in an attempt to gather additional information on the reported event; however, no additional information was obtained.However, if additional information becomes available, this report will be supplemented accordingly.This reported event has been reported by the importer on mdr# 2951238-2020-00313.
 
Event Description
On december 19, 2019, the service center received medwatch report # (b)(4) that states, "patient was undergoing a cystoscopy, right retrograde pyelogram, right ureteroscopy, laser lithotripsy procedure.As the surgeon proceeded to remove the ureteral scope from the right ureter it became lodged in the mid to distal ureter.The surgeon could not free the scope without significant traction.Despite attempts, the scope remained impacted within the ureter.A 14 french foley was replaced into the bladder, and ureteral scope was secured to patient's right thigh.The patient was then transferred to tertiary facility for further evaluation and management.".
 
Manufacturer Narrative
This report is being updated to provide additional information reported by the patient's attorney (reported in b5, b7), and corrected information (h6-health effect clinical code).
 
Event Description
Additional information received: the patient was hospitalized for a period of 4 days and the scope was able to be removed by the surgeon without surgical intervention.
 
Manufacturer Narrative
This supplemental report was submitted to provide additional information from the legal manufacturer (lm) and to update the following sections: g3, g6, h2, h4, h6 and h10.The lm performed the device history records for the concerned device and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A review of the repair history indicated that there was no repair history within the past year.A review of the complaint history indicates there are no similar complaints and no similar failures from the user facility.The investigation was completed by the lm and determined that there is no manufacturing, material or processing related cause for this failure mode.The concerned scope was not returned for evaluation; therefore, the exact cause of the reported malfunction could not be conclusively determined.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key9578354
MDR Text Key189485954
Report Number8010047-2020-00873
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170286988
UDI-Public04953170286988
Combination Product (y/n)N
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberURF-P5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2019
Initial Date FDA Received01/10/2020
Supplement Dates Manufacturer Received07/09/2021
09/07/2021
Supplement Dates FDA Received08/03/2021
09/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
Patient Weight79
-
-