• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS PRESTIGE CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS PRESTIGE CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Catalog Number 6973060
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2019
Event Type  malfunction  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient underwent cervical disc arthroplasty at c6-7 due to degenerative disk disease(ddd).Intra op, inserter did not retain the implant properly when impacted with the mallet.There were no patient complications as a result of this event.
 
Manufacturer Narrative
Product analysis results: the upper (superior) and lower (inferior) tips of the inserter shaft have been bent to where one side at the.4mm dimension located on the print at location f6 on page 1 of 2 appear to be touching.This is consistent with over-torque if information is provided in the future, a supplemental report will be issued.
 
Event Description
Mdr decision corrected to not reportable.No additional supplemental reports will be required unless additional information received indicates reportable event.
 
Manufacturer Narrative
H1: review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunction.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE CERVICAL DISC SYSTEM
Type of Device
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key9579882
MDR Text Key183786713
Report Number1030489-2020-00053
Device Sequence Number1
Product Code MJO
UDI-Device Identifier00613994714848
UDI-Public00613994714848
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6973060
Device Lot NumberEM17G012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2019
Initial Date FDA Received01/13/2020
Supplement Dates Manufacturer Received03/25/2020
12/19/2019
Supplement Dates FDA Received04/07/2020
02/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-