• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION AORTIC AP; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION AORTIC AP; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 505DA20
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Ischemia (1942); Occlusion (1984); No Information (3190)
Event Date 12/17/2019
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that immediately post implant of this aortic mechanical valve, myocardial ischemia resulting from coronary artery occlusion was noted.The patient's vital signs became unstable, and the valve was explanted and replaced with a non-medtronic valve.Vital signs improved and no additional adverse patient effects were reported.
 
Manufacturer Narrative
Medtronic received additional information that the 20mm mechanical valve was replaced with a 19mm non-medtronic mechanical valve.No additional adverse patient effects were reported.  h6: coding corrected medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC AP
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
3800 annapolis lane
minneapolis MN 55447
MDR Report Key9580273
MDR Text Key174669189
Report Number3008592544-2020-00005
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
PMA/PMN Number
P990046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/17/2023
Device Model Number505DA20
Device Catalogue Number505DA20
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2019
Initial Date FDA Received01/13/2020
Supplement Dates Manufacturer Received01/15/2020
Supplement Dates FDA Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient Weight67
-
-