Model Number 102096-67A |
Device Problem
Material Separation (1562)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/13/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
|
Event Description
|
On (b)(6) 2019 senseonics was made aware of an event where the sensor broke into two pieces before the removal procedure.Both pieces of the sensor was removed during the initial removal procedure.
|
|
Manufacturer Narrative
|
Manufacturers narrative is updated to: all fragments of the broken sensor were removed on (b)(6) 2019.The patient was doing fine afterwards.The definitive root cause could not be determined for the sensor breakage.Based on the case notes, it is possible that the patient sustained an impact to the insertion site even though no trauma was reported.It is also possible that the hcp unknowingly broke the sensor during the removal procedure.Spontaneous breakage is exceedingly unlikely.
|
|
Manufacturer Narrative
|
Both the pieces of sensor were broken and removed prior to removal on (b)(6) 2019.The patient was doing fine.No further investigation is required.H6 results code updated to 3221.H6 conclusion code updated to 4311.
|
|
Search Alerts/Recalls
|