Age at time of event: 18 years or older.Device evaluated by manufacturer: the device was returned for analysis.The returned complaint device consisted of a rotablator rotalink burr catheter.The detached burr was not returned.The advancer, handshake connections, sheath and coil were microscopically and visually examined.The coil is stretched and broken 135.5cm from the strain relief.Inspection of the remainder of the device presented no other damage or irregularities.
|
Reportable based on device analysis completed on 23dec2019.It was reported that the handshake connector broke away.The target lesion area was located in the left anterior descending artery.A 1.50mm rotalink burr was selected for use.During preparation, it was noted that the burr catheter broke away while connecting it to the advancer at the handshake connection.The procedure was completed with another of the same device.No patient complications were reported, and the patient was stable.However, the device analysis revealed that the coil was broken 135.5cm from the strain relief.
|