• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAZOR ROBOTICS LTD MAZOR X; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAZOR ROBOTICS LTD MAZOR X; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number TPL0059
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2019
Event Type  malfunction  
Manufacturer Narrative
Export files have been received however, analysis has not yet been completed.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that there was an inaccurate screw placement.During a t9 down to l2 percutaneous case, they did two registrations because they split the ct up.The first was t9 to l1, then the second went down to l2.The first registration looked good but when they moved down to l2, the left screw was placed medial and the right screw was also placed medial (in the pedical, but medial).Troubleshooting involved recalibrating the c-arm.The issue extended the surgical time by less than 1 hour.The amount of deviation was unknown.
 
Manufacturer Narrative
H2) additional information: a1-a5) patient information provided.H3) after eliminating sub-optimal planning, patient or platform shift, and registration was verified to be accurate, the investigating team has reviewed all available information and concluded the most probable cause of the deviation of bilateral l2 trajectories sub-optimal surgical technique, which have resulted in medially deviated trajectories, as the tools were most probably subjected to soft-tissue pressure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from a manufacturer representative reported that the screws were deviated between 3.5 and 10 mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAZOR X
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il CO 30795 67
IL  3079567
MDR Report Key9586661
MDR Text Key175393395
Report Number1723170-2020-00140
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K163221
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Initial Date Manufacturer Received 12/17/2019
Initial Date FDA Received01/14/2020
Supplement Dates Manufacturer Received01/20/2020
03/03/2020
Supplement Dates FDA Received02/13/2020
03/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age60 YR
Patient Weight60
-
-