OBERDORF SYNTHES PRODUKTIONS GMBH PFNA-II BLADE L90 TAN; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Catalog Number 04.027.053S
Device Problems Difficult to Remove (1528); Material Protrusion/Extrusion (2979)
Patient Problems Pain (1994); No Code Available (3191)
Event Date 12/19/2019
Event Type  Injury  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Pma/510k: device is not distributed in the united states, but is similar to device marketed in the usa.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2019 the patient underwent surgery with the proximal femoral nail antirotation (pfna), the incision was closed with the blade protruding (reported by (b)(4)).On (b)(6) 2019 the patient underwent removal surgery.The surgeon tried another removal method which had not been performed in the first surgery but the blade could not be removed.The surgeon cut the blade with a cutting tool and the surgery was completed successfully with more than 30 minutes delay.The removal surgery could relieve the patient¿s pain.There was no plan to perform the revision surgery.No further information is available.Concomitant devices: unknown end caps ( part# unknown, lot# unknown, quantity 1), unknown locking screw ( part# unknown, lot# unknown, quantity unknown).This report is for one pfna-ii nail.This is report 2 of 2 for (b)(4).This report is linked to (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history part: 04.027.053s, lot: 5l85102, manufacturing site: bettlach, release to warehouse date: 30.Aug.2019, expiry date: 01.Aug.2029.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFNA-II BLADE L90 TAN
Type of Device
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9587936
MDR Text Key175089112
Report Number8030965-2020-00364
Device Sequence Number1
Product Code HSB
UDI-Device Identifier07611819394223
UDI-Public(01)07611819394223
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.027.053S
Device Lot Number5L85102
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/19/2019
Initial Date FDA Received01/14/2020
Supplement Dates Manufacturer Received01/16/2020
Supplement Dates FDA Received02/06/2020
Patient Sequence Number1
Treatment
UNK - END CAPS: PFNA-II; UNK - SCREWS: LOCKING
Patient Outcome(s) Required Intervention;
Patient Age80 YR
-
-