Model Number 10445 |
Device Problems
Entrapment of Device (1212); Difficult to Remove (1528); Material Twisted/Bent (2981); Adverse Event Without Identified Device or Use Problem (2993); Physical Resistance/Sticking (4012)
|
Patient Problem
Vascular Dissection (3160)
|
Event Date 12/19/2019 |
Event Type
Injury
|
Event Description
|
It was reported that a dissection occurred.A 14f isleeve introducer sheath was inserted for a valve replacement procedure.Balloon angioplasty was performed.During removal, the balloon catheter became stuck at the tip of the isleeve.The tip of the isleeve became bunched up.The isleeve was exchanged for another of the same device and the procedure was completed.A dissection was noted but the patient continued to have good flow and no treatment was performed.
|
|
Event Description
|
It was reported that a dissection occurred.A 14f isleeve introducer sheath was inserted for a valve replacement procedure.Balloon angioplasty was performed.During removal, the balloon catheter became stuck at the tip of the isleeve.The tip of the isleeve became bunched up.The isleeve was exchanged for another of the same device and the procedure was completed.A dissection was noted but the patient continued to have good flow and no treatment was performed.
|
|
Manufacturer Narrative
|
Returned product consisted of an isleeve device that had a non-bsc balloon catheter stuck in the lumen with a non-bsc inflation device connected to the hub of the balloon catheter.Analysis confirmed kinks throughout the isleeve and the seams expanded.The balloon was bunched up on the distal end.Functional testing was attempted to remove the device, however, due to the damage of the devices removal was not possible.The photos were reviewed and it was confirmed that there are numerous sheath kinks/buckles along the sheath.At lease one of the seams is starting to expand as expected with device usage.The damaged bav device is seen inside of the isleeve.
|
|
Search Alerts/Recalls
|