Model Number 427
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
The stent has not been returned to numed so the complaint can't be confirmed.Additional information has been requested, but nothing has been received.All covered cp stents are inspected for proper covering attachment in final inspection.A sample from each lot is tested for covering attachment strength.The sample from this lot of stents had a covering attachment strength of 3.24lbs, which is well above the minimum specification.The production traveler (device history record) was reviewed and no issues were found.All devices in this lot met the criteria for release and distribution.There are no other associated complaints with this lot of devices.
 
Event Description
As per the report from the user facility / foreign distributor - the expandable sleeve of eptfe fell off the stent as the physician tried to mount it on a balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERED CP STENT
Type of Device
AORTIC STENT
Manufacturer (Section D)
NUMED, INC.
2880 main street
hopkinton NY 12965
Manufacturer (Section G)
NUMED, INC.
2880 main street
hopkinton NY 12965
Manufacturer Contact
nichelle laflesh
2880 main street
hopkinton, NY 12965
3153284491
MDR Report Key9592553
MDR Text Key196068429
Report Number1318694-2020-00002
Device Sequence Number1
Product Code PNF
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
P150028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date11/30/2022
Device Model Number427
Device Catalogue NumberCVRDCP8Z39
Device Lot NumberCCP-0823
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/17/2019
Initial Date FDA Received01/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-