• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER Back to Search Results
Catalog Number SBI07008013P
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/13/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information: a guide catheter was not used.When the balloon passed through the introducer, the balloon fractured.No detachment occurred.The balloon was removed from the patient intact without intervention.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to use an in.Pact admiral paclitaxel eluting balloon catheter during a procedure to treat a mildly calcified lesion in the proximal iliac artery.The vessel diameter and lesion length are 8mm and 60mm respectively.The vessel was moderately tortuous and mild calcification is reported.The device was inspected and negative prep performed with no issues.The lesion was pre dilated.The device passed through a previously-deployed stent.It was reported that during advancement through the guide catheter, a break/fracture occurred on the catheter and the detached portion was removed from patient by removing the balloon.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT ADMIRAL PACLITAXEL-ELUTING PTA BALLOON CATHETER
Type of Device
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key9595519
MDR Text Key188425125
Report Number9612164-2020-00264
Device Sequence Number1
Product Code ONU
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P140010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/13/2020
Device Catalogue NumberSBI07008013P
Device Lot Number0008922398
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2020
Initial Date FDA Received01/16/2020
Supplement Dates Manufacturer Received01/22/2020
Supplement Dates FDA Received01/27/2020
Date Device Manufactured12/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-