• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER Back to Search Results
Model Number DR8064
Device Problems Peeled/Delaminated (1454); Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
For the reported event, lot number was provided, and lot history review.The sample was returned for evaluation.Balloon rupture, and peeling pebax was confirmed for the device.A root cause has not been determined.The device was labeled for single use.
 
Event Description
This report summarizes one malfunction.A review of the reported information indicated that model dr8064.Pta balloon dilatation catheter allegedly experienced material rupture and peeled.This report was received from a single source.This event did involve patient with no reported patient injury.Age, weight, and gender were not provided for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Type of Device
DORADO PTA DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens TX 75751
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key9597314
MDR Text Key178117870
Report Number2020394-2020-00517
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00801741058721
UDI-Public(01)00801741058721
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDR8064
Device Catalogue NumberDR8064
Device Lot Number93ND0098
Initial Date Manufacturer Received 12/31/2019
Initial Date FDA Received01/16/2020
Type of Device Usage N
Patient Sequence Number1
-
-