Inspection of photos of the affected device showed that the device had been inadequately "repaired" by unauthorized third parties who did not adhere to the original manufacturing specifications.The etched label was covered with tape and the originally matte-finished instrument was sandblasted.Silver solder was used on the jaw of the instrument, as indicated by the yellowish color of the tip.This material is not used during the manufacturing process and is not adequate for this device.To complete the investigation, the device itself and the broken-off insert will need to be inspected.Return of the device has been requested.
|
While using the 121140 carb-bite mayo-hegar nh 7, the tungsten carbide insert came off the needle holder.Piece was retrieved and sent with needle holder to clinical engineering.There was patient contact, however, there was no patient injury or harm.There was no surgical delay.Both the device and the broken piece were saved.
|