Model Number 27050CA
Device Problem Material Separation (1562)
Patient Problems Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
Event Type  malfunction  
Manufacturer Narrative
Per the factory, upon examination we found that the ceramic tip is broken off.The ceramic insert has not been provided.There was sufficient glue between tip and sheath, because the bonded ceramic still adheres well in the inner part of the sheath.The sheath shows proximal corrosion, which can be attributed to incorrect cleaning.Damages like this happen if the ceramic tip is hit against hard objects or edges.The damage of the product is not caused by a production problem or material defect.Hints on how to inspect the ceramic tip are already included in the ifu.
 
Event Description
Per the factory in (b)(4), allegedly, there was an event occurred in (b)(6) that the tip was broken off during an operation.A second operation was needed to remove the broken part.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH
Type of Device
UROLOGY RESECTOSCOPE INNER SHEATH
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm,
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm,
Manufacturer Contact
susie chen
2151 e. grand avenue
el segundo, CA 90245
4242188201
MDR Report Key9603154
MDR Text Key194177713
Report Number9610617-2020-00004
Device Sequence Number1
Product Code FBO
UDI-Device Identifier04048551077412
UDI-Public4048551077412
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K943668
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27050CA
Device Catalogue Number27050C
Device Lot NumberNU01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/12/2019
Initial Date FDA Received01/17/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-