• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED; FOLEY CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED; FOLEY CATHETERS Back to Search Results
Model Number 123518A
Device Problems Burst Container or Vessel (1074); Use of Device Problem (1670)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the balloon burst during use of the catheter.No pieces were observed to be missing, and the balloon was filled with 4 1/2cc.No medical intervention was required.
 
Event Description
It was reported that the balloon burst during use of the catheter.No pieces were observed to be missing, and the balloon was filled with 4 1/2cc.No medical intervention was required.
 
Manufacturer Narrative
The reported event was confirmed as use-related.An evaluation was conducted on the returned catheter and based on the evaluation, an irregular burst was noted on the balloon sac.The user misused the catheter and did not use the device according to the ifu provided and also used vaseline to lubricate the catheter prior to insertion.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿bardia foley catheter caution: this product contains natural rubber latex which may cause allergic reactions.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile unless package is opened or damaged warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer syringe.Do not use needle.Visually inspect the product for any imperfections or surface deterioration prior to use.Do not use if package is opened or damaged.Recommended inflation capacities 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.¿ h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER SILICONE COATED
Type of Device
FOLEY CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9604939
MDR Text Key180399443
Report Number1018233-2020-00420
Device Sequence Number1
Product Code EZC
UDI-Device Identifier00801741095108
UDI-Public(01)00801741095108
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/28/2023
Device Model Number123518A
Device Catalogue Number123518A
Device Lot NumberMYDR4212
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2020
Initial Date Manufacturer Received 12/26/2019
Initial Date FDA Received01/19/2020
Supplement Dates Manufacturer Received02/19/2020
Supplement Dates FDA Received02/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-