This report is being submitted as follow up no.1 to update section h3, and to provide the completed investigation results.The actual sample was received for evaluation.Visual inspection of the two actual samples upon receipt revealed no obvious anomaly, such as a break, in their appearance.Both actual samples were rinsed and subjected to another closer visual inspection and confirmed not to have any anomaly, such as a break, which could have led to the clogging of the fiber.Each actual sample was built into a circuit with tubes.With a help of a roller pump, bovine blood (ht 25%, 37°c) was let to flow through the circuit.As a result, normal flow rate was obtained in both samples.Both actual samples were tested for their ultrafiltration performance.The obtained results from both were confirmed to meet the manufacturer specifications.Both actual samples were tested for their pressure loss during the circulation at the flow rate of 500ml/min and tmp of 400mmhg.The obtained results from both were confirmed to meet the manufacturer specifications.Ifu states: adequate heparinization of the blood is required in order to prevent it from clotting in the system.The capiox hemoconcentrator is designed to operate at flow rates within the range of 100 to 500 ml/min during ultrafiltration.Do not use blood flow rates outside this range.Less than 100 ml/min blood flow may cause blood coagulation in the device.Do not exceed 50% hematocrit at the blood outlet during ultrafiltration.Do not stop blood flow during extracorporeal circulation as it may cause clotting in the system.A blood flow of least 50 ml/min is recommended even if ultrafiltration is stopped by clamping the filtrate line.Bubbles in the system may cause clotting and blood damage.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.The investigation results verified the returned sample was of the normal product.It is likely that due to some factor(s), the rate of the flow going into the actual sample was decreased, causing blood to get concentrated excessively and plug the fibers; blood which had become high in concentration due to the administration of packed red blood cells or due to removal of water contained in blood entered the actual sample, got further concentrated and plugged the fibers.However, the exact cause of the reported event cannot be definitively determined based on the available information.
|