It was reported that while using the slap hammer extractor handle, the handles threads became flat and would not screw back into slap hammer attachment.While using the thin osteotome handle holder, the dial to tighten the thin osteotome in place became loose and the threads were broken as well.We believe both these instruments have been used so much that the lifespan of the instruments broke because regular wear and tear over time.These instruments had no negative affect on the outcome of the case and no pieces broke off that could have caused a problem.No surgical delay.
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|