• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ORIGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D003
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Displays Incorrect Message (2591)
Patient Problem No Code Available (3191)
Event Date 12/02/2019
Event Type  Injury  
Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis.(b)(4).
 
Event Description
It was reported that this implantable cardioverter defibrillator (icd) recorded code 1005 indicative shorted shock on the right ventricular (rv) lead.Intervention was performed.The device was explanted and the right ventricular (rv) lead was capped.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Patient code 3191 captures the reportable event of additional surgical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN MINI ICD DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9611917
MDR Text Key175712656
Report Number2124215-2019-28377
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526533761
UDI-Public00802526533761
Combination Product (y/n)N
PMA/PMN Number
P960040/S306
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/17/2019
Device Model NumberD003
Device Catalogue NumberD003
Device Lot Number248343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2020
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received01/21/2020
Supplement Dates Manufacturer Received01/29/2020
Supplement Dates FDA Received04/06/2020
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
Patient Age52 YR
-
-